NOTE:
MEDRAD Avanta Table mount: A 76.2 mm (3 in) and max 25 m (984 in) length conduit
between technical room and patient room shall be prepared below the floor for the three
injector cables. It is recommended to use the MEDRAD Avanta floor mounting bracket
to cover the duct hole in the patient room if there is no gases box.
Floor mount installation can be accomplished one of two ways:
Connectors mounted in trough under mounting bracket (Figure 1)
Connectors mounted above mounting bracket (Figure 2)
For further MEDRAD Avanta floor mounting, see the Installation guide
Avanta Floor Mounting Bracket.
178
Innova 2121-IQ/3131-IQ Biplane Cardiovascular Imaging System Pre-Installation Manual
Illustration 6-17: MEDRAD Avanta floor mounting methods
Direction 5177951-4-1EN, Revision 1
MEDRAD
17 Cable Channeling