Fluoro Ups Option - GE Innova 2121-IQ Biplane Preinstallation Manual

Cardiovascular imaging system
Table of Contents

Advertisement

GE Healthcare
Room
Core System
Note (3): For UPS 20 kVA option refer to
Note (4): On batteries heat output = 0.52 kW / 1755 BTU
Note (IVUS): For IVUS environmental limitations and heat output, refer to

2.2.2 Fluoro UPS Option

The heat produced by the UPS is transferred to the environment by its ventilation. Cooling air
enters the cabinets through the air inlet (grids) located at the bottom and exhausted through the
outlet on the roof. A suitable ventilation or cooling system must be installed to extract the heat from
the UPS room.
If the UPS is placed on a raised floor, the airflow for UPS cooling should enter from underneath
the UPS, through the appropriate aperture on the raised floor.
If the UPS runs in a dusty environment, we recommend strongly to install filters on the air inlet of
the UPS room. In this case it should be considered that these filters can cause reduced speed at
the air inlet.
The size of the air inlet has therefore to be dimensioned accordingly.
Contact your Local Distributor or one of the Service Centre, which will help you to find valuable
solutions.
58
Innova 2121-IQ/3131-IQ Biplane Cardiovascular Imaging System Pre-Installation Manual
HEAT OUTPUT
Stand by
kW
Section 2.2.2
CAUTION
Make sure there is a ventilation air flow, preferably ensured by natural air flow,
otherwise by enforced ventilation, so that hydrogen concentration is below 1%
(according to Standard IEC 62040-1-2).
NOTICE
Do not put anything on the top of the cabinet.
Moderate Use (8 cas‐
es / a 10 hour day)
BTU/hr
kW
BTU/hr
Section 2.2.3
Illustration 3-1:
2 Environmental Requirements/Limitations
Direction 5177951-4-1EN, Revision 1
Maximum Use (maxi‐
Typical Use (13 cas‐
mum peak power dur‐
es / a 10 hour day)
ing exam)
kW
BTU/hr
kW
BTU/hr