GE Healthcare
Chapter 3 Physical Requirements of room
1 Presentation of the 3 Rooms
1.1 Examination room
•
Innova LC Positioner
•
Innova LP4 Positioner
•
Omega Patient Table
•
Performix X–Ray Tube Assembly (x 2)
•
Collimator (x 2)
•
Innova Digital Detector (x 2)
1.2 Technical room
•
C2 LC/LP Cabinet
•
C1 Frontal Cabinet and C1 Lateral Cabinet
•
Coolix 4000 Chiller (2 Cabinets)
•
Detector Conditioner HEAT DRY 1 (2 Cabinets)
•
Fluoro UPS (option) (1 Cabinet)
•
Power Distribution Box (Main Disconnet Panel)
•
3 kVA cabinets UPS
1.3 Control room
•
VCIM
•
DL Flat Panel
•
DL Keyboard
•
Monitors
Chapter 3 Physical Requirements of room
Innova 2121-IQ/3131-IQ Biplane Cardiovascular Imaging System Pre-Installation Manual
WARNING
THE ELECTRONIC CABINETS (C1 FRONTAL, C1 LATERAL, C2, COOLIX 4000
CHILLERS, DETECTOR CONDITIONERS AND FLUORO UPS WHEN
INSTALLED) INCLUDE FANS THAT ARE CREATING AIR–CIRCULATION OF
PULSED–AIR. WHEN THIS PULSED AIR IS IN AN ENVIRONMENT THAT MAY
CONTAIN AIRBORNE PATHOGENS LIKE AN EXAM ROOM/CONTROL ROOM,
THERE IS A RISK OF TRANSMISSION OF THESE AIRBORNE PATHOGENS
FROM PATIENTS TO OTHER PATIENTS OR CLINICAL PERSONNEL
(NOSOCOMIAL DISEASES).
TO REDUCE THIS RISK, THE ELECTRONIC CABINETS MUST BE INSTALLED
IN A ROOM SEPARATED FROM EXAM ROOM/CONTROL ROOM, I.E.,
TECHNICAL ROOM.
Direction 5177951-4-1EN, Revision 1