NOTICE
Installation safety instructions:
Contractor responsibility:
Electrical contractor is responsible for providing and connecting the cables
and configuring the PDB in by-pass mode.
GEHC is responsible for powering on the system with the UPS in by-pass
mode.
GEDE is responsible for UPS comissioning.
After removing the sidewalls of the UPS, make sure that all earth connections
when reassembling, are correctly reattached.
This UPS is intended for use in a controlled indoor environment free of
conductive contaminants and protected against animals intrusion.
HIGH GROUND LEAKAGE CURRENT: Ground connection is essential before
connecting to AC input!
For Europe only, if a differential breaker is placed on the hospital main supply,
upstream the PDB, the differential shall be set to 300 mA.
Switching OFF the unit does not isolate the UPS from the utility.
Do not install the UPS in an excessively humid environment or near water.
Avoid spilling liquids on or dropping any foreign object into the UPS.
The unit must be placed in a sufficiently ventilated area; the ambient
temperature should not exceed 104°F (40°C).
Optimal battery life time is obtained if the ambient temperature does not exceed
77°F (25°C).
It is important that air can move freely around and through the unit. Do not block
the air vents.
Avoid locations in direct sunlight or near heat sources.
Check local regulations for UPS installation.
NOTICE
Storage safety instructions:
Store the UPS in a dry location; storage temperature must be within –13°F (–
25°C) to 131°F (55°C).
If the unit is stored for a period exceeding 3 months, the battery must be
recharged periodically (time depending on storage temperature).
NOTICE
Battery safety instructions:
The battery–voltage is dangerous for person's safety.
Never dispose of battery in a fire: They may explode.
Do not open or mutilate battery: Their contents (electrolyte) may be extremely
toxic.
If exposed to electrolyte, wash immediately with plenty of water.
Avoid charging in a sealed container.