Electrical Requirements; Ethernet Network Requirement - GE Innova 2121-IQ Biplane Preinstallation Manual

Cardiovascular imaging system
Table of Contents

Advertisement

GE Healthcare
NOTE:
Ensure sufficient cable length on the Procedures Room end of the cables according to the
table below
PIM cable
Control Station cable
Joystick cable
ECG cable
NOTE:
The incorrect cable length might cause reliability issues during the operation !
Make sure, that the out of conduit parts of the cables with the connectors are properly protected
during the rest of the installation

5.4 Electrical Requirements

The power for the IVUS hardware components delivered with the product (PC, monitor, Video
Switch and printer) is supplied through the IVUS Safety Isolation Transformer .
The customer is in charge to provide a wall outlet to supply the Safety Isolation Transformer. This
outlet shall be located in the control room suitable to the IVUS components positioning. The
maximum rated power of the Safety Isolation Transformer is 600 VA, the maximum continuous
system power consumption for s5i and s5iGE IVUS systems is 400 VA.
The s5i and s5iGE IVUS systems can be configured to work on 100, 120, or 230 VAC. The
difference between the supported configurations is limited switching the fuse values. See table
below :
Frequency: 50/60 Hertz
The country specific power cords are pre-selectable in the GTC.

5.5 Ethernet Network Requirement

The IVUS CPU located in the control room shall be connected to the hospital Ethernet network. A
wall Ethernet outlet shall be available in the control room to connect the device.
Chapter 6 Electrical Connections
Innova 2121-IQ/3131-IQ Biplane Cardiovascular Imaging System Pre-Installation Manual
Same applies to the 2nd Ethernet cable, if a later extension of the IVUS remote
controls expected.
Cable
Line Voltage
100 VAC
120 VAC
230 VAC
NOTICE
The IVUS will be connected to a standard hospital outlet. There is no grounding
requirement specified at site preparation.
Direction 5177951-4-1EN, Revision 1
Free length from the conduit exit
2.4 m (8 ')
2.1 m (7 ')
2.7 m (9 ' )
Depends on the ECG location
Fuse value
8 A
6.3 A
3.15 A
161