Patient Environment Equipment; Recommended Room Lighting - GE Innova 2121-IQ Biplane Preinstallation Manual

Cardiovascular imaging system
Table of Contents

Advertisement

GE Healthcare
3kVA UPS FOR CE CONFIGURATION WITHOUT FLUORO UPS), including UPS and Fluoro
UPS outputs.
To remove input power from THE INPUT CABLE OF THE 3kVA UPS (FOR CE CONFIGURATION
WITHOUT FLUORO UPS), turn OFF the PDB main breaker
Protect the Emergency Stop from accidental actuation.

1.5 Patient Environment Equipment

The components that may be installed within patient vicinity need to be medical equipment ("patient
vicinity is defined in the standardization as a space within the room 1.83 m (6 ft) beyond the
perimeter of the examination table and extending vertically 2.29 m (7 ft, 6 in) above the floor.").
For the Innova System, the equipment are:
Table
C–arm
Monitors
Injector

1.6 Recommended Room Lighting

Zone 1: Intensive lighting needed by operator. Manually cut not located at false ceiling level.
86
Innova 2121-IQ/3131-IQ Biplane Cardiovascular Imaging System Pre-Installation Manual
Illustration 4-1:
Direction 5177951-4-1EN, Revision 1
1 Room Layout Considerations