11 Door Interlocks

Chapter 6 Electrical Connections
Innova 2121-IQ/3131-IQ Biplane Cardiovascular Imaging System Pre-Installation Manual
CAUTION
IEC 60601–2–43 requires not to install door interlocks. It is the responsibility of the
field service to check that this requirement is not in contradiction with local
regulation. In case of conflict, follow local regulation. No other measures employed
for radiation protection should cause the interruption of irradiation and any other
disturbance of a procedure in progress.
Illustration 6-16: C2 cabinet bulkhead RM DOOR connecting
Direction 5177951-4-1EN, Revision 1
171