GE Innova 2121-IQ Biplane Preinstallation Manual page 90

Cardiovascular imaging system
Table of Contents

Advertisement

GE Healthcare
Location
Exam room
For exam and control rooms, several configurations are possible.
The pre–installation instructions shall define a room layout where the location of the remote
controls versus the moving parts of equipment. This layout shall define the maximum distance
between the remote control location and equipment and the axis of the equipment (L axis) versus
the remote controls (dead angles concern).
NOTE:
Motion controls installed in remote location from the table shall be installed at a location
where all the positioner axis are visible by the operator.
Refer to Illustrations in
90
Innova 2121-IQ/3131-IQ Biplane Cardiovascular Imaging System Pre-Installation Manual
Constraint Name
minimum rail length past
iso-centre towards foot end
WARNING
CARRIAGE COVERS CAN ENCLOSE DUST PARTICLES. CARE SHOULD BE
TAKEN TO AVOID PROPAGATION.
IT IS RECOMMENDED TO AVOID DIRECT AIR FLOW BETWEEN LP RAILS.
Room Layout Drawings
Parameter
Specification
Length
1.3 m (4 ft, 7 in)
to see possible exam / control rooms layouts.
Direction 5177951-4-1EN, Revision 1
Comment
LP4 parking at patient foot end not
allowed
1 Room Layout Considerations