GE Healthcare
•
Establish a 76 mm (3") conduit below floor from the IVUS CPU location in the Control Room,
to the foot end area of the Patient Table in the Procedures Room:
○
Locate the Procedures Room exit of the conduit close to the centerline of the Patient Table
○
Locate it as close to the Patient Table base as possible, considering :
■
Room for access to the table for service purposes
■
Limitation from other cables connected to the table
■
Location of the table plate
■
Location of Mac-Lab conduit
■
Location of Physio or Med Gases connections
■
The IVUS cables are moving cables, they have to allow the Patient Table motions
(Panning, Lifting and Rotating)
•
Use similar solutions (i.e. protection against cleaning fluids, etc. ) as it is used for the Mac-Lab
conduit
•
See also guidance on Typical Layout drawing 4-58f sheets E1 and S2, showing Mac-Lab and
P/M Gas location recommendations.
5.3 Run Interconnection cables through the Conduit
Use the IVUS PIM (#806365003) as reference to the procedures below.
•
Unpack , and check the completeness of the Preinstallation kit :
○
1pcs - Custom PIM cable 30 m (98 ft, 5 in) - Volcano ID :806452001
○
2pcs - CAT5 cable 30 m (98 ft, 5 in) - Volcano ID :803391001
○
1pcs - Custom Twin axial cable - Volcano ID :806380001
○
1pcs – Ground Cable, 30 m (98 ft, 5 in) s5i – Volcano ID :806889001
○
1pcs - IVUS preinstallation Manual - Volcano ID :806365003
•
Install 806889001 Ground Cable through the Control Booth conduit (see note)
•
Install the cables through the IVUS conduit from the Control Room end :
○
806452001 Room - PIM cable, round shaped connector end to the Procedures
○
803391001 - Ethernet cable (1pcs)
•
Install the optional cables through the IVUS conduit from the Control Room end :
○
803391001 - Ethernet cable if 2 IVUS remote control items will be installed 1 IVUS Control
Station and 1 IVUS Joystick)
○
806380001 - ECG cable in case, if NON Mac-Lab ECG will be installed
NOTE:
NOTE:
NOTE:
160
Innova 2121-IQ/3131-IQ Biplane Cardiovascular Imaging System Pre-Installation Manual
For the ground cable, see the IVUS service manual for special instructions
concerning cable running and cutting .
If the installation has the MaC-Lab ECG in the configuration, the ECG cable shall
not be run through the conduit, it will be used in the Control Room to connect
Mac-Lab and IVUS PCs.
You should consider running the ECG cable through the conduit in case, the
customer requires interconnecting to a NON Mac-Lab ECG at the table base.
In this case you shall hide the unconnected cable in the conduit for future
expansion options.
Direction 5177951-4-1EN, Revision 1
5 IVUS Rev 1 Option