3.3.5 Water Pipe Requirements

Chapter 3 Physical Requirements of room
Innova 2121-IQ/3131-IQ Biplane Cardiovascular Imaging System Pre-Installation Manual
Illustration 3-6
NOTICE
Due to the plastic bushing used in the USA to protect cables from the sharp edges
of conduits it is necessary to place the cable conduit inside the table cable access
opening but the height of the outcoming conduit plus bushing is limited to 12 mm
(1/2 in).
Chemical anchors Pull out efforts and recommendations in
for pull out effort on each fixation bolts.
Illustration 3-7: Water Conduit location
NOTICE
In some countries, depending on local regulations, it may be forbidden to run
electrical cables and water pipes in the same conduit. In this case, two separate
conduits are required.
Direction 5177951-4-1EN, Revision 1
ANGIO / CARDIO
NA
1278 mm (50.3 in)
1278 mm (50.3 in)
CARDIO / NEURO
1395 mm (54.9 in)
1395 mm (54.9 in)
1395 mm (54.9 in)
Chapter 5,
71