2.4.6 Installations Requirements & Environment Control
In order to minimize interference risks, the following requirements shall apply:
NOTE:
In order to avoid interferences, the same AC power distribution panel should
supply all components, accessories, the Innova system (& subsystems as the
Advantage Workstation). The separated AC power line should supply the panel.
56
Cables shielding & grounding:
All interconnect cables to peripheral devices must be shielded and properly
grounded. Use of cables not properly shielded and grounded may result in
the equipment causing radio frequency interference.
Separated Power supply distribution panel & separated power line:
This product complies with the radiated emission limits as per the
CISPR11 Group1 ClassA standard.
The Innova system is predominantly intended for use (e.g. in hospitals)
with a dedicated supply system, and with an X–ray shielded room.
In case of using in a domestic environment (e.g. doctors' offices), in
order to avoid interferences, it is recommended to use a separated AC
power distribution panel & separated power line, and an X–ray shielded
room.
Subsystem & accessories Power supply distribution:
All components, accessories subsystems, systems which are electrically
connected to the Innova system, have to be all AC power supplied by the
same power distribution panel & line.
Stacked components & equipment:
The Innova system should not be used adjacent to or stacked with other
equipment; if adjacent or stacked use is necessary, the Innova system
should be observed to verify normal operation in the configuration in which
it will be used.
Low frequency magnetic field:
In case of an Innova system, the Gantry (digital detector) shall be apart 1
meter from the generator cabinet, and 1 meter apart from the analog (CRT)
monitors. These distances specifications shall minimize the low frequency
magnetic field interference risk.
Static Magnetic Field Limits : The Innova system is compatible with the
Earth magnetic Field. Earth magnetic Field is lower than 1 Gauss.
Electrostatic discharges environment & recommendations:
In order to reduce electrostatic discharge interference, install a charge
dissipative floor material to avoid electrostatic charge buildup.
The relative humidity shall be at least 30 percent.
The dissipative material shall be connected to the system ground
reference, if applicable.
Direction 5314528-2-1EN, Revision 1