GE Healthcare
Table 3-4
indicates the typical heat dissipation at nominal load and specify cooling air flow in two
situations:
•
1000 m (3280 ft) altitude, cooling air between 25°C (77°F) and 30°C(86°F)
•
250 m (820 ft) altitude, cooling air below 18°C (64°F)
Table 3-4:
Fluoro UPS
CE version
UL version
2.2.3 IVUS Option
2.2.3.1 Environmental Limitations in Use (room climate)
Table 3-5: IVUS Rev 1
Humidity (non condensing) :
Temperature (at 20% Relative Humidity ) :
Altitude (Relative to see level) :
Table 3-6: IVUS Rev 2
Humidity (non condensing) :
Temperature (at 20% Relative Humidity ) :
Altitude (Relative to see level) :
2.2.3.2 Heat Output
PC
Monitor
Printer
2.3 IEC60601–1–2 Electromagnetic Standard Compliance & Documentation
2.3.1 General Scope
This equipment complies with IEC-60601-1-2: Edition 2.1 and Edition 3 EMC standard for medical
devices.
The Innova system is suitable to be used in the electromagnetic environment, as per the limits &
recommendations described in the tables here after:
•
Emission Compliance level & limits
Chapter 3 Physical Requirements of room
Innova 2121-IQ/3131-IQ Biplane Cardiovascular Imaging System Pre-Installation Manual
Typical heat output
1.50 kW (5101 BTU/hr)
4720 BTU/hr (1.38 kW)
Device
Control room
Control room
Control room
CAUTION
Take any steps necessary to provide correct temperature and dust free
environment for correct operation of PC in the control room.
On–line cooling air flow at 30°C /
86°F
625 m 3 /h (368 CFM)
301 CFM (511 m 3 /h)
30 - 80 %
15 - 32 C
max 3048 m
30 - 80 %
5 - 35 C
max 3048 m
Location
(Table
3-7).
Direction 5177951-4-1EN, Revision 1
Typical on-line cooling air flow at
18°C / 64°F
227 m 3 /h (134 CFM)
119 CFM (203 m 3 /h)
(60 - 90 F)
(max 10000 ft)
(41 - 95 F)
(max 10000 ft)
Heat output
335 W – 1206 BTU/h
125 W – 427 BTU/h
67 W – 229 BTU/h (idle)