GE Innova 2121-IQ Biplane Preinstallation Manual page 158

Cardiovascular imaging system
Table of Contents

Advertisement

GE Healthcare
Table 6-3: System FEEDER from hospital
Feeder run length:
Feeder run length:
Recommended feeder cable section in mm² : for PHASES, NEUTRAL & GROUND PDB input terminals accept 35² to 95².
Bigger cables needs a size reduction before PDB entry.
NOTE:
Increasing transformer power will decrease its voltage loss and feeder cables section. To be calculated site by site.
158
Innova 2121-IQ/3131-IQ Biplane Cardiovascular Imaging System Pre-Installation Manual
Power supply voltage 480 V
Panel (PDB) to C1 run in m (ft)
8 (26)
16 (52)
10 m (30 ft)
2/0
20 m (70 ft)
2/0
30 m (100 ft)
2/0
40 m (130 ft)
2/0
50 m (160 ft)
2/0
60 m (200 ft)
3/0
80 m (260 ft)
4/0
100 m (330 ft)
300
120 m (390 ft)
400
8 m
10 m
35
20 m
35
30 m
50
40 m
50
50 m
70
60 m
95
80 m
120
100 m
185
120 m
240
NOTICE
24 (79)
2/0
2/0
2/0
2/0
2/0
3/0
2/0
250
3/0
300
4/0
400
350
600
500
see Note
600
see Note
Power supply voltage
380 V
Panel (PDB) to C1 cabinet run
16 m
24 m
8 m
35
35
35
35
70
35
70
95
35
70
120
50
95
185
70
120
240
70
185
300
95
240
see note
150
300
see note
185
4 Power and Grounding Recommendations
Direction 5177951-4-1EN, Revision 1
WARNING
RECOMMENDED FEEDER CABLE GAUGE :
FOR PHASES, NEUTRAL & GROUND PDB
INPUT TERMINALS ACCEPT 2/0 AND 3/0
AWG. BIGGER CABLES NEEDS A SIZE RE‐
DUCTION BEFORE PDB ENTRY.
400 V
16 m
24 m
8 m
35
35
35
35
50
35
50
70
35
70
95
50
95
120
70
95
185
70
150
300
95
240
see note
120
300
see note
185
415 V
16 m
24 m
35
35
35
50
50
70
70
95
95
120
95
150
120
240
185
see note
240
see note