10 Room Speaker

The remote speaker provides the operator in the exam room, the indication of record exposure
sequence (same level information as provided by the control room control panel).
The speaker must be installed in the exam room out of the patient vicinity to give easy access to
the FE for maintenance operations and to medical staff to adjust the tone signal level and to be
able to see the power LED indicator.
Patient Vicinity Definition : see
Considerations
170
Innova 2121-IQ/3131-IQ Biplane Cardiovascular Imaging System Pre-Installation Manual
WARNING
THE REMOTE SPEAKER KIT SHALL NOT BE INSTALLED IN THE PATIENT
VICINITY.
Patient Environment Equipment in
Illustration 6-15:
Direction 5177951-4-1EN, Revision 1
Chapter 4, Room Layout
10 Room Speaker