GE Innova 2121-IQ Biplane Preinstallation Manual page 146

Cardiovascular imaging system
Table of Contents

Advertisement

GE Healthcare
Cable
Recommended (1)
R rectifier
5 x 10 mm²
L out /load
5 x 10 mm²
(1) Size and type may be adapted locally per local regulation.
146
Innova 2121-IQ/3131-IQ Biplane Cardiovascular Imaging System Pre-Installation Manual
Illustration 6-2: Power distribution - UL
CE UPS configuration
Type
3ph+N+PE
3ph+N+PE
UL UPS configuration
Max gauge
Recommended (1)
AWG3
AWG6
AWG3
AWG6
Direction 5177951-4-1EN, Revision 1
Type
3ph+N+PE
3ph+N+PE
1 Power Distribution
Max length
24 m (78ft 8in)
24 m (78ft 8in)