GE Innova 2121-IQ Biplane Preinstallation Manual page 120

Cardiovascular imaging system
Table of Contents

Advertisement

GE Healthcare
The UPS cabinet can be positioned against the wall but, in order to improve the ventilation and to
make easier the maintenance operations for UPS and battery, we recommend a minimum distance
of 200 mm (8") from the wall.
The right side of the UPS cabinet must be accessible for maintenance operation.
In front of the cabinet, the clear width of the service area to insure electrical safety shall be at least
0.9 m. In cases where 2 cabinets are installed face to face (both sides of the access way), the
clear width shall be at least 1.2 m.
Recommended minimum clearance between ceiling and top of the UPS should be 400 mm (16")
for proper cooling air exhaust.
120
Innova 2121-IQ/3131-IQ Biplane Cardiovascular Imaging System Pre-Installation Manual
Illustration 5-22: Fluoro UPS Cabinets (Optional) - UPS CE
Direction 5177951-4-1EN, Revision 1
1 Dimension Drawings

Advertisement

Table of Contents
loading

This manual is also suitable for:

Innova 3131-iq biplane

Table of Contents