NOTE:
Prepare the floor such that the Tilt Table baseplate will be flush with the floor surface.
For alternative table bolts or seismic area, refer to template drawing shown in Illustration
Gantry and table mounting holes contained in
3.3.4 Hole dimension and preferred location in concrete floor
In the examination room, the Innova LC Positioner is not placed on a computer floor but directly
put on concrete floor, the location of the cable access needs to be carefully planned.
Otherwise, if the cable run is located under the concrete floor, the cables will have to come through
the floor and in this case you will need two holes, one for the LC Positioner and the other for the
patient table.
The diameter of both holes must be the same 225 mm or 9 in.
Chapter 3 Physical Requirements of room
Innova 2121-IQ/3131-IQ Biplane Cardiovascular Imaging System Pre-Installation Manual
Direction 5177951-4-1EN, Revision 1
Chapter 5, Mounting
Requirements.