Injectors; Remote Injector (Rack Mount); Pedestal Injector; Injector L.f. Illumina - GE Innova 2121-IQ Biplane Preinstallation Manual

Cardiovascular imaging system
Table of Contents

Advertisement

GE Healthcare

13 Injectors

The injector is provided with an additional ground cable (P/N #2135737) to meet equipotentiality
requirement at patient vicinity.
Injector emergency power off:
if no emergency power off is required for the injector, the injector can be powered from a wall
outlet socket.
if emergency power off circuit must include the injector, ensure the power for the injector is via
the PDB.

13.1 Remote Injector (rack mount)

A power cord is supplied with the injector.
110 VAC or 230 VAC are supplied from the main disconnect room device. The injector must be
connected to this supply.

13.2 Pedestal Injector

A power cord is supplied with the injector.
In all cases (110 VAC or 230 VAC), it will be connected to a wall outlet near the operator location,
fed by the main disconnect room device.

13.3 Injector L.F. ILLUMINA

As a result of power demand availability for all functions of this new injector, in any cases it will be
connected to a wall outlet sized to 1500 W, 110 VAC or 230 VAC as nominal voltage. The wall
outlet is fed by the main disconnect room device.
Chapter 6 Electrical Connections
Innova 2121-IQ/3131-IQ Biplane Cardiovascular Imaging System Pre-Installation Manual
Direction 5177951-4-1EN, Revision 1
173

Advertisement

Table of Contents
loading

This manual is also suitable for:

Innova 3131-iq biplane

Table of Contents