GE Healthcare
Access to a working phone at the facility
7
for emergency use, including MR magnet
delivery.
All walls primed (final coat not needed on
Day 1), and counter tops that will support
8
equipment must be installed. No dust-pro‐
ducing cabinetry work in installation areas.
Mechanical supplier has been provided
with a set of equipment installation draw‐
9
ings for reference. For California, permit‐
ted construction drawings or PMI-speci‐
fied installation drawings are required.
Conduit/electrical cable ducting/dividers/
access flooring installed, with the excep‐
tion of surface-mounted floor ducting. Wir‐
10
ing to the main disconnect panel is
installed and compliant with equipment in‐
stallation drawings or pre-installation man‐
ual.
Issued Date: 7/9/07 Rev 11
Chapter 7 Additional Planning Aids
Innova 2121-IQ/3131-IQ Biplane Cardiovascular Imaging System Pre-Installation Manual
GEHC Only: COE # (888) 799.7266 Option 5 (PMI Support)
X
X
X
X
X
X
X
X
Direction 5177951-4-1EN, Revision 1
X
X
203