Calibration Screen; Calibration Set-Up - Agilent Technologies Helium User Manual

Leak detector
Table of Contents

Advertisement

Technical Information

Calibration Screen

Calibration Set-Up

Press Calibration Menu screen.
The Calibration screen appears (see the following figure). Use this screen to:
The calibration routine can only be executed while in Test/FineTest.
As of February 2023 Firmware Rev LD01.06
120822: Show Home Cal - Selection.
The Home Screen Calibration button and
Calibration Status Pop Up Messages can be
enabled from the Calibration Menu Screen.
Figure 28 Calibration Set-Up
Factory default – Show Home Cal & Show Cal Pop (not selected/not enabled).
NOTE
Agilent Helium Leak Detector User Manual
View the value and expiration date of the Internal Calibrated Leak
Enter the External Calibrated leak value
Select whether to use an Internal or external calibrated leak
Set the calibration mode: Full or Fast Calibration