Advertisement

Technical Information
Table 24 Soft power shutdown routine via RS232 (turbo breaking)
Command
SOFT_S
initiates turbo breaking shutdown procedure
?SOFTS_C
Table 25 Calibration Required at Startup (Flashing "C")
Command
1 FORCE_CALPW
0 FORCE_CALPW
? FORCE_CALPW
Table 26
Startup Calibration Messages – Cancel of Execute Calibration at startup (20-minute bypass)
Command
S_CALIBRATE
1 DISMISS_20M
2 DISMISS_20M
1 DISMISS_10R
156
Action
Response – Wait for 2 minutes
Response – Shut down complete
0 (not initiated)
1 (in progress)
2 (complete)
Action
Calibration Required at Startup
Shut off cal required at Startup
0 (not initiated)
1 (initiated)
Action
Execute start up calibration message
Execute No/Cancel calibration request
Execute calibration/Yes
10
Cancels E
range external calibration request
Agilent Helium Leak Detector User Manual