The commands listed in Table A-3 are used to cause an immediate change in the
spectrometer operating parameters. These Commands Do Not Change The Non-
Volatile Operating Parameters.
Table 18 Spectrometer Operating Parameters
Command
PUT-EXPONENT
PUT-ION
PUT-RANGESTOP
The commands listed in the following table are used to cause certain leak detection
actions.
Table 19 Leak Detection Actions
Command
AUTO
CALIBRATE
DECREMENT
DISABLE-0ENABLE
DISABLE-APP
DISABLE-PASSWORD
DISABLE-xREJECT
DISABLE-RANGESTOP
DISABLE-VOLUME
ENABLE-0ENABLE
ENABLE-PASSWORD
Agilent Helium Leak Detector User Manual
Parameter
Sets the exponent leak range for Manual mode. Preceded by a two-character number
Acceptable values are 0 through -10.
Sets the ion value of the ion source. Preceded by a three-digit number of volts, in the range
200 to 350.
Sets the Range Stop exponent. Preceded by two or three characters -XX or -X.
Parameter
Initiates auto-ranging mode. Success is indicated by the normal ok response.
Initiates a Full or Fast calibration depending on system settings. The CPU software tunes,
then adjusts the gain so that the current helium signal causes the current leak rate
measurement to be the same as the most recently input using INIT-STDLEAK.
Subtracts 1 from the Manual Ranging leak rate exponent (making it more negative).
Success is indicated by the normal ok response. Does nothing in auto-ranging mode. Does
nothing upon reaching the most sensitive range.
Turns the Zero Enable function OFF.
Disables the active Applications Setup.
Turns the password function OFF.
Turns the Reject Set Point OFF. Where x is a number from 1 to 4 or the letter A.
Turns the Range Stop function OFF.
Turns audio OFF.
Turns the Zero Enable function ON.
Turns the Password feature ON.