Table 17 Non-Volatile Operating Parameters (Continued)
Command
INIT-ION
INIT-LANGUAGE
INIT-LINEAR
INIT-LKEXPIRE
INIT-LEAKTEMP
INIT-1LOG
INIT-ML-XFER
INIT-PASSWORD
X INIT-PWDLEVEL
INIT-QUICK-CAL
INIT-RANGESTOP
INIT-1REJECT
INIT-2REJECT
INIT-3REJECT
Agilent Helium Leak Detector User Manual
Parameter
Sets the ion voltage of the ion source. Preceded by a three-digit number (volts) in the range
220 to 280.
Selects the language to be shown on the display. Preceded by 0-6 where (0=English,
1=French, 2=German, 3=Korean, 4=Spanish, 5=Japanese, 6=Chinese).
Sets the leak rate analog output voltage to Linear. Not preceded by a value.
Sets the date on which the internal calibrated leak expires. Preceded by the date.
Example: mm dd yyyy INIT-LKEXPIRE.
Sets the internal calibrated leak temperature (printed on the leak label) that the factory
calibration was performed at. Preceded by 12.0 to 50.0 (degrees C).
Example: 23.5 INIT-LEAKTEMP
Sets the leak rate analog output voltage to logarithmic at 1 volt per decade. Not preceded
by a value.
Sets the value of the Fine Test-flow mode crossover pressure in Torr. Preceded by X.XE-XX.
The range is 1.0E-03 to 1.0E-01 Torr. The default is 1.0E-01 (100 mTorr).
Sets the system password. Preceded by a 1 to 6 digit password and followed by a digit
indicating the number of digits in the password. Example: 123123 6 INIT-PASSWORD.
Password is read back using the ?SYSPWD command.
X is the password level from 1 to 3.
Sets which type of calibration will be run if the CALIBRATE command is initiated. Preceded
by 0 or 1, 0 for FULL cal, 1 for FAST cal.
Sets the value of the most sensitive range exponent. Preceded by XX.
Sets the leak rate value for Reject Set Point #1. Preceded by a two-digit leak rate mantissa
with a decimal point after the first digit followed by E-, and then by a two-digit leak range
exponent, the helium leak rate number is in atm cc/sec.
Example 3.4E-08 INIT-1REJECT
Same as INIT-1REJECT, but for Reject Set Point #2.
Same as INIT-1REJECT, but for Reject Set Point #3.