The rear panel of the leak detector (Model G8610A and G8610B) is shown in the
following figure.
Figure 22 Rear Panel (with Discrete I/O PCB)
I/O Connector is a 25-pin female D-sub connector to interface with an external PLC
via a parallel cable.
Inputs:
 Four level inputs: optically isolated, (+5 to+24) VDC.
 Four momentary inputs: optically isolated, (+5 to+24) VDC, requires 200 ms
All inputs are opto-isolator LEDs with a series 3300 Ohm resistor.
 Outputs: Eight optically isolated emitter followers with a 10 Ohm series
 Power: An external power supply is supplied by the user.
Agilent Helium Leak Detector User Manual
7
9
pulse width.
resistor, 14 mA max drive current, +24VDC max.