Technical Information
The password can also be read through the RS-232 port in the event of a lost
NOTE
password. See Section A "Communications Protocol".
Lock Vent Control
Lock All Control

Wireless

Information relating to wireless communications
MAC Address
SYNC Channel
SS ID
Channel
Agilent Helium Leak Detector User Manual
Disables the VENT button. This option prevents
accidental system venting during test while under
remote or PLC control. The Vent Lock feature is not
active when the Auto Sequencer is on.
Disables the Test, Vent and Zero buttons. This option
prevents accidental system venting, testing, system
hold and Zeroing during test while under remote or PLC
control.
Two MAC addresses appear on this screen:
The MAC address at the top of the screen is the leak
detector MAC address.
The MAC address above the hand held button is the
MAC address of the hand held device.
For informational and factory debug purposes only.
Displays the SS ID number, factory set to 086, for
informational and factory debug purposes only.
Click the button and a keypad appears to configure the
channel (1-19) that the unit uses to communicate with the
wireless remote. Up to 10 wireless remotes and leak
detectors can be communicating in a common area at a
time.
121