Screen Settings

The Screen Settings is a personalized menu that configures the leak detector's bar
graph, leak rate ranging and the analog output voltage type.
Figure 61 Settings: Screen Settings
Buttons and Fields
Bar Graph
Agilent Helium Leak Detector User Manual
Configures the bar graph:
Linear Scale - The linear bar graph leak rate appears
across the top of all screens when selected. In the Linear
Bar Graph display mode, the bar graph display represents
the mantissa of the leak rate, spanning from 0 to 10. The
numerical display (located to the upper right of the bar
graph) indicates the leak rate exponent value. This mode
offers exceptional resolution within a decade.
Log Scale - The log bar graph leak rate appears across
the top of all screens. In Log Bar Graph display mode, the
bar graph display represents the entire leak rate range of
the leak detector, spanning from E-11 atm cc/sec to E-4
atm cc/sec. The numerical display (located to the upper
right of the bar graph) is not lit in this mode.
115