Technical Information
Audio Set Point
The Audio Setpoint is very useful in the Sniffing Application. The programmed Audio
NOTE
Set Point value can be programmed 50% < Set Point 1, which alarms the operator
that they are approaching a leak (leak location).
Plugged Probe Alarm
Files
Calibrate
88
Desired Audio Alarmed Leak Rate Value (> <). This
value can be adjusted from 10
Point triggers with its' own distinct tone if different
from Set Point 1.
A distinct high frequency tone different from the Audio
& Set Point.
1
This alarm is useful in alerting the operator that the
desired Power Probe/Sniffing Probe adjustment
(pressure) has changed. A High Limit can be
selected in alerting that too much flow (high test
port pressure) is present. A Low Limit trigger is
typically the norm indicating that the Power
Probe/Sniffing is blocked or the tubing is kinked or
pinched off.
The operator can store up to three setup
configurations:
•
Sniff 1-3
The information icon provides the necessary directions
for a successful external calibration.
External calibration/Sensitivity check: Follow the step
by step process.
Agilent Helium Leak Detector User Manual
-8
to 10
-4
. The Audio Set