Navigate to the Manual Tuning Menu and:
5
Navigate to the Calibration Menu and turn the Read Standard Leak – OFF.
Agilent Helium Leak Detector User Manual
Set the GAIN to 1.0.
Set the EMISSION to between 0.600 and 0.6500.
Adjust the ION value up or down to attain the maximum value. Ideally, the
ION value should be between 205 and 240.
Adjust the EMISSION to attain the maximum value. Ideally, the EMISSION
value should be between 0.600 and 1.200.
Set the GAIN so that the leak rate on the screen matches the value of the
internal leak you wrote down.