Agilent Technologies Helium User Manual page 155

Leak detector
Table of Contents

Advertisement

Technical Information
Table 20 Baud Rate Setting
Baud Rate Setting Input
New RS232 Commands as of February 2023 (Rev. LD01.06)
Table 21
Command
XBAUDR
?BAUDR
Table 22
Command
?FIL1
?FIL2
DISMISS_NOEMISS
?READY
?CAL_S (calibration routine status)
Table 23 Disable & Enable Touch Screen Display
Command
1 LOCK_S
0 LOCK_S
? LOCK_S
DISMISS_CAL OK
Agilent Helium Leak Detector User Manual
0
1
2
3
4
Action
(input 0 thru 4) X = 0 to 4
(return 0 thru 4)
Action
0 Filament burnt out
1 Filament good/not active
2 Filament good/ active
=ok Filament No Emission Message
Response Yes (System Ready)
Response No (System Ready)
0 not initiated
1 in progress
2 complete
Action
Locks out the touch screen display
Unlocks the touch screen display/normal operation
Return: 0 (touchscreen not locked)
Return :1 (touchscreen locked)
Automatically selects OK for both "Cal OK" and "Cal Failed" pop ups
Baud Rate
9600
19200
38400
57600
115200
155