Installation Requirements - Agilent Technologies Helium User Manual

Leak detector
Table of Contents

Advertisement

Instructions for Use

Installation requirements

Installation requirements are described in the following table.
Table 1
Installation Requirements
Item
Leak Detector Location
Power
Mains Supply Voltage Fluctuations
Pollution
Fixed Mains Socket Outlet
32
• Close to its power source.
• Two to three feet clearance for proper ventilation.
• Sufficient room for the operator to safely maneuver to perform testing.
NOTE: Additional ventilation requirements for your specific application are
discussed in Section "Recommended Additional Services".
NOTE: When installing a bench system with a mechanical pump, the power
cord for that pump, if connected to the back of the leak detector, cannot exceed
three meters. Additionally, connect the mechanical pump to the leak detector
using the shortest possible vacuum connection to maximize vacuum
performance.
See Table 5.
Must not exceed ±10% of the nominal voltage.
Pollution Degree 2 per EN61010-1 Standard.
Connect to a building protective earthing system.
Agilent Helium Leak Detector User Manual