Maintenance: Gauge Calibration - Agilent Technologies Helium User Manual

Leak detector
Table of Contents

Advertisement

Technical Information
Fore Pump Type and Exhaust Pump Activation
The fore pump type and exhaust pump are configured at the factory or can be
configured by a service engineer.
When the Last Service date is updated/changed the Operating Hours reset to Zero.
NOTE

Maintenance: Gauge Calibration

A display of <1.0E-03 indicates the pressure is less than the lowest displayable
pressure of the TC gauge, which is 1.0E-03. Use this screen to perform a calibration
on the test port thermocouple gauge. Calibrate the thermocouple gauge for the
most accurate operation.
Select Gauge Calibration from the Maintenance Menu selection.
Use this screen to:
Set a calibration point at atmosphere.
Set a calibration point at a low vacuum level (<= 1E-4 Torr).
Figure 77 Maintenance: Gauge Calibration
134
Agilent Helium Leak Detector User Manual