Test Port Vacuum
Test Port ATM
Test Port Pressure and
Spectrometer Pressure
Test Port Pressure Gauge Calibration Procedure
Calibrates the test port pressure thermocouple gauge. Perform both the Vacuum
and Atmospheric Calibration procedures in the order given.
Maintenance: System Defaults
Use this screen to reset Turbo hours plus Filaments 1and 2 hours and restore
system defaults.
This screen displays the on/time hours for the turbo pump and filaments.
When replacing a turbo pump or a filament ensure that the hours are reset. This
aids in scheduling an effective maintenance service.
Figure 78 Maintenance: System Defaults
Agilent Helium Leak Detector User Manual
Calibrates the thermocouple gauge zero value. Do
this at or below 1E-4 Torr.
Sets the thermocouple atmospheric value. Perform
this operation only when the test port is at
atmosphere
These values appear on this screen for reference.