Technical Information
For Plugged Probe Alarm - High Pressure Test (Sniffing Mode) see Section "Sniff
NOTE
Method".
Off:
The audio sounds for all leak rates above zero.
Set points may be turned ON or OFF (enabled or disabled). When the Audio Set
Point is OFF, audio sounds for all leak rates above zero. When the audio set
point is ON, audio sounds only for leak rates above the audio set point.
To configure reject set points and the audio set point:
1
Press the blue check for the Reject Set Point.
2
Press the value button associated with the set point. The data entry pad
appears.
3
Enter the set point and press OK.
4
Press the < or > soft key to select whether the reject setpoint activates above or
below the programmed value.
5
Press the blue check for the Audio Set Point.
6
Press the value button associated with the audio set point. The data entry pad
appears.
7
Enter the set point and press OK.
Agilent Helium Leak Detector User Manual