Password and Restrictions

The Password and Restrictions Menu is an optional configuration that protects the setup
of the leak detector. The factory and operational setups of the leak detector significantly
affect the performance of the leak detector and must only be accessed by trained service
personnel. This menu also allows protection from unintentional calibrations and
accidental operator input while the machine is unattended or under test.
Figure 65 Systems Settings: Password and Restrictions
VENT, TEST and ZERO buttons are still accessible in all password levels.
NOTE
Buttons and Fields
Restrictions
Set Password
120
Activates/deactivates the password function:
Medium: Allows operation (Home Screen) of the Test,
Vent & Zero functions. Locks out all menu selections
other than the Calibration Menu. Allows Internal &
External Calibration.
High:Locks out all menu selections (Home screen only.
Changes the password. Enter up to six numeric values for the
password and confirm the password selection. An empty
entry is not allowed. If no password is entered, the old
password is used. The default password from the factory is
777777. You need this when changing to a new password.
Agilent Helium Leak Detector User Manual