Calibration Values Invalid - Agilent Technologies 1100 Series Reference Manual

Diode array and multiple wavelength detectors
Hide thumbs Also See for 1100 Series:
Table of Contents

Advertisement

3
Troubleshooting and Test Functions

Calibration Values Invalid

The calibration values read from the spectrometer ROM are invalid.
After recalibration, the calibration values are stored in ROM. The processor
periodically checks if the calibration data are valid. If the data are invalid or
cannot be read from the spectrometer ROM, the error message is generated.
• Defective detector main board.
• Defective optical unit.
70
Probable Causes
Suggested Actions
Exchange the detector main board.
Exchange the optical unit.
1100 Series DAD and MWD Reference Manual