Agilent Technologies Helium User Manual page 125

Leak detector
Table of Contents

Advertisement

Technical Information
Figure 70 Internal Calibrated Leak Keypads
2
Enter the required values as shown on the calibrated leak label or calibration
certification.
3
Repeat step 1 and 2 for each desired item.
4
Press OK to exit the screen.
5
The value of the internal calibrated leak, as shown on the CALIBRATION Menu,
is compensated for temperature using the calibrated leak temperature sensor.
The calibrated leak is not compensated for temperature if any of the following
conditions exist:
Agilent Helium Leak Detector User Manual
The temperature sensor on the calibrated leak fails or there is a cable
connection problem.
The temperature sensor reading is out of the operating range. The
operating temperature range is from 10.0 to 45.0° C (50 to 113° F).
The internal calibrated leak is not a E-7 or E-8 range leak.
125