Cover Violation - Agilent Technologies Infinity Binary Pump VL 1290 User Manual

Table of Contents

Advertisement

7
Error Information
General Error Messages

Cover Violation

Error ID: 7461
The top foam has been removed.
The sensor on the main board detects when the top foam is in place. If the
foam is removed while the lamps are on (or if an attempt is made to switch on
for example the lamps with the foam removed), the lamps are switched off,
and the error message is generated.
Probable cause
1
2
84
The top foam was removed during
operation.
Foam not activating the sensor.
Suggested actions
Please contact your Agilent service
representative.
Please contact your Agilent service
representative.
Agilent 1290 Infinity Binary Pump User Manual