Error Information
Pump Error Messages

Current of primary pump drive too high

Error ID: 22433
The current of the primary pump drive is too high.
• Parameter: 1 or 4 referring to pump drive.
Probable cause
1
2

Current of secondary pump drive too high

Error ID: 22434
The current of the secondary pump drive is too high.
• Parameter: 2 or 3 referring to pump drive
Probable cause
1
2
96
Blockage of flow path between primary
pump head and pressure sensor, e.g. of the
heat exchanger.
Primary pump drive is defect.
Blockage of flow path between secondary
pump head and pressure sensor, e.g. of the
heat exchanger.
Secondary pump drive is defect.
Suggested actions
• Check for blockages in flow path.
• Please contact your Agilent service
representative.
Please contact your Agilent service
representative.
Suggested actions
• Check for blockages in the flow path.
• Please contact your Agilent service
representative.
Please contact your Agilent service
representative.