Unknown purge valve type
Error ID: 22435
The type information of the purge valve is invalid.
• Parameter: None
Probable cause
1
2
Pump drive encoder error
Error ID: 22437
The pump drive encoder has generated no signal.
• Parameter: 1 – 4 referring to pump drive
Probable cause
1
Pump drive error
Error ID: 22438, 22439
The pump drive failed during calibration.
• Parameter: 1 – 4 referring to pump drive
Probable cause
1
Agilent 1290 Infinity Binary Pump User Manual
Wrong valve head installed.
Valve head has invalid RFID tag content.
Pump drive encoder is defect.
Pump drive motor defect or mechanically
blocked.
Error Information
Pump Error Messages
Suggested actions
Check or replace purge valve head.
Check or replace purge valve head.
Suggested actions
Please contact your Agilent service
representative.
Suggested actions
Please contact your Agilent service
representative.