3
Install new Jet Weaver or flip the Jet Weaver for
backside.
Agilent 1290 Infinity Binary Pump User Manual
Changing configuration or replacing the Jet Weaver
4
Reinstall the capillary connections.
The inlet at the bottom of the Jet Weaver is connected to
the central port of the pump valve by a capillary (length
300 mm, 0.17 mm i.d.). The outlet at the top is connected
to the autosampler.