Maintenance
Replacing Parts of the High Pressure Filter Assembly
1
Remove the capillary connection from the outlet of the
secondary pump head.
3
Replace the filter frit and seal cap as desired. Please note
the correct orientation of the filter frit.
158
2
Use a 14 mm hex wrench for opening the filter assembly
of the secondary pump head
4
Re-install the filter assembly using the torque wrench
(14 mm hex bit) set to 16 Nm.
Agilent 1290 Infinity Binary Pump User Manual