Replacing the Solvent Selection Valve (SSV)

When
In case of problems with the solvent selection valve
Parts required
#
1
4
1
Close shut-off valve. Remove tubing connections
between the SSV and the solvent shut-off valves and the
SSV and the degassing unit inlets.
Agilent 1290 Infinity Binary Pump User Manual
p/n
Description
G4280-60029
Solvent selection valve
G4220-60035
Tubing kit 140 mm SSV to shutoff valve or degassing unit (2 tubes)

Replacing the Solvent Selection Valve (SSV)

2
Push down the SSV panel for removing it.
9
Maintenance
125