Drive current too high

Error ID: 22409
The current consumption of the pump drive is too high.
• Parameter: 1 – 4 referring to pump drive
Probable cause
1
2

Drive timeout

Error ID: 22410
Drive is blocked mechanically, fails during initialization.
• Parameter: 1 – 4 referring to pump drive
Probable cause
1

Overcurrent of pump drive

Error ID: 22411
The current consumption of the pump drive is too high.
• Parameter: 1 – 4 referring to pump drive
Probable cause
1
2
Agilent 1290 Infinity Binary Pump User Manual
Blockage of system before pressure sensor.
Drive motor defect.
Blockage of pump drive Drive motor defect.
Blockage of system before pressure sensor.
Drive motor defect.
Error Information
Pump Error Messages
Suggested actions
Check for blockage of e.g. outlet valve filter frit,
purge valve, heat exchanger.
Please contact your Agilent service
representative.
Suggested actions
Please contact your Agilent service
representative.
Suggested actions
Check for blockage of e.g. outlet valve filter frit,
purge valve, heat exchanger.
Please contact your Agilent service
representative.