Leak

Error ID: 0064
A leak was detected in the module.
The signals from the two temperature sensors (leak sensor and
board-mounted temperature-compensation sensor) are used by the leak
algorithm to determine whether a leak is present. When a leak occurs, the leak
sensor is cooled by the solvent. This changes the resistance of the leak sensor
which is sensed by the leak-sensor circuit on the main board.
Probable cause
1
2

Open Cover

Error ID: 0205
The top foam has been removed.
Probable cause
1
2
Agilent 1290 Infinity Binary Pump User Manual
Loose fittings.
Broken capillary.
Foam not activating the sensor.
Dirty or defective sensor.
Error Information
General Error Messages
Suggested actions
Ensure all fittings are tight.
Exchange defective capillaries.
Suggested actions
Please contact your Agilent service
representative.
Please contact your Agilent service
representative.