Setting the 8-bit Configuration Switch

Special Settings

The special settings are required for specific actions (normally in a service
case).
The tables include both settings for modules – with on-board LAN and without on-board
N O T E
LAN. They are identified as LAN and no LAN.
Boot-Resident
Firmware update procedures may require this mode in case of firmware
loading errors (main firmware part).
If you use the following switch settings and power the instrument up again,
the instrument firmware stays in the resident mode. It is not operable as a
module. It only uses basic functions of the operating system for example, for
communication. In this mode the main firmware can be loaded (using update
utilities).
Table 13
Boot Resident Settings (On-board LAN)
Mode Select
TEST/BOOT
218
SW1
SW2
SW3
1
1
SW4
SW5
1
0
0
Agilent 1290 Infinity Binary Pump User Manual
SW6
SW7
SW8
0
0
0