Table 5
Method parameters
Parameter
Limits
Pressure Limits
Max: 1200 bar
(17400 psi ) for flow
rates up to 2 mL/min
. For flow rates
between 2 mL/min
and 5 mL/min , the
maximum pressure
ramps down to
800 bar (11600 psi).
Min: any value
between 0 and the
upper pressure limit
setting.
Timetable
Advanced
External Contacts
Agilent 1290 Infinity Binary Pump User Manual
Setting up the Pump with the Instrument Control Interface
Description
Sets the maximum and minimum pressure limits for the pump.
• Max is the maximum pressure limit at which the pump will switch
itself off, protecting the analytical system against over-pressure.
• Min is the minimum limit at which the pump will switch itself off, for
example, if any solvent reservoir is empty, this prevents system
damage by pumping air.
See
"Timetable Settings"
See
"Advanced Settings"
The External Contacts section enables you to set up the switching of the
external contacts.
N O T E
The External Contacts section is present only when a
BCD/external contacts board is installed.
on page 53
on page 52
4
Using the Pump
51