Agilent Technologies Infinity Binary Pump VL 1290 User Manual page 152

Table of Contents

Advertisement

9
Maintenance
Assembling the Pump Head
15
For the secondary pump head, assemble and install the
high pressure filter assembly using the torque wrench
(14 mm hex wrench) set to 16 Nm.
16 Nm
17
Insert the pump piston.
Piston
152
16
Remove the alignment piston.
18
Insert both pump heads to the link plate and make sure
that the clips snap in that fix the pump heads.