Automatic Configuration With Bootp; About Agilent Bootp Service - Agilent Technologies Infinity Binary Pump VL 1290 User Manual

Table of Contents

Advertisement

Automatic configuration with Bootp

All examples shown in this chapter will not work in your environment. You need your own
N O T E
IP-, Subnet-Mask- and Gateway addresses.
Assure that the detector configuration switch is set properly. The setting should be either
N O T E
BootP or BootP & Store, see
Assure that the detector connected to the network is powered off.
N O T E
If the Agilent BootP Service program is not already installed on your PC, then install it from
N O T E
your Agilent ChemStation DVD, located in folder BootP.

About Agilent BootP Service

The Agilent BootP Service is used to assign the LAN Interface with an IP
address.
The Agilent BootP Service is provided on the ChemStation DVD. The Agilent
BootP Service is installed on a server or PC on the LAN to provide central
administration of IP addresses for Agilent instruments on a LAN. The BootP
service must be running TCP/IP network protocol and cannot run a DHCP
server.
Agilent 1290 Infinity Binary Pump User Manual

Automatic configuration with Bootp

Table 16
on page 228.
13
LAN Configuration
237

Advertisement

Table of Contents
loading

Table of Contents