Preparing the Pump Module for Transport

When
If the module shall be transported or shipped.
Parts required
#
1
1
1
Preparations
Flush both solvent channels with isopropanol.
Mechanical damage
C A U T I O N
1 Remove solvent inlets from solvent reservoirs. Disconnect the solvent
Agilent 1290 Infinity Binary Pump User Manual
p/n
9301-0411
9301-1337
G4220-44000
For shipping the module, insert the Protective Foam to protected the module from
mechanical damage.
Be careful not to damage tubing or capillary connections while inserting the module
in the Protective Foam.
tubing from the inlet of primary pump heads for both solvent channels. Use
a syringe for removing liquid from the solvent tubings between solvent
reservoir, shutoff valve panel, solvent selection valve, degassing unit and
pump inlets. Switch the solvent selection valve if applicable.

Preparing the Pump Module for Transport

Description
Syringe; Plastic
Syringe adapter
Protective Foam
9
Maintenance
163