Changing configuration or replacing the Jet Weaver
1
Remove capillary connections from the Jet Weaver.
128
2
Remove the hex screws that fix the Jet Weaver to the
pump housing.
N O T E
The standard Jet Weaver (Jet Weaver 35 µL/100 µL
(G4220-60006)) has a front and a rear side with
different internal volumes (35 / 100 µL) that are
optimized for a low delay volume or best mixing
performance. Please refer to recommendations in the
Agilent 1290 Infinity System Manual.
The optional Jet Weaver (Jet Weaver 380 µL
(G4220-60012)) is recommended for applications
which are challenging with respect to mixing noise
(e.g. TFA applications) and has just one side.
Agilent 1290 Infinity Binary Pump User Manual